Electrochemical cell produced by DD Scientific for the detection of Carbon Monoxide, with high sensibility to Hydrogen.
Measurement Range 0 – 1000 ppm
Maximum Overload 2000 ppm
Linearity: within ± 5%
Repeatability <± 5%
Response time, T90 <30 s
Output Signal 50 ± 20 nA / ppm
Long-term Output Drift <5% per annum
Recommended Load Resistor 10 ohms
Resolution (Electronics dependent) < 1 ppm typical