Nominal Range: 0-2000ppm
Maximum Overload: 4000ppm
Output Signal: 0.075±0.025μA/ppm
Typical Baseline Range (pure air): -2 to +17ppm equivalent
Bias Voltage: 20 or +250mV
Recommended Gain: 0.5 to 2.5
Weight: 26g
Inboard Filter: To remove acid gases/alcohol
Auxiliary Electrode: To compensate for H2 crossinterference
Expected Operating Life: 3 years
Resolution: 1ppm
Temperature Range: -20°C to +50 °C
Operating Range: 800-1200mBar
Storage Range: 800-1200mBar
Max Diff (capillary to amb.): ±100mBar
Pressure Coefficient: 0.02% signal/mBar
T90 Response Time: <40 seconds
Relative Humidity Range: 15% to 90% non-condensing
Long Term Output Drift: <10% signal loss/year
Repeatability: 2% of signal
Output Linearity: Linear