PERFORMANCE CHARACTERISTICS Nominal Range: 0-40,000ppm Operating Principle: 4-electrode electrochemical Measurement Range: 0-1000 ppm NO Maximum Overload: 0-100,000ppm Inboard Filter: To remove acid gases Filter Life: > 400,000 ppm hours (1000ppm NO @ 200ml/min) Expected Operating Life: Three years in air Output Signal: 0.010 ± 0.004 ?A/ppm Resolution: 10ppm Temperature Range: -20°C to +50°C Pressure Range: Atmospheric ± 10% Pressure Coefficient: 0.007 ± 0.003 %signal/mBar T90 Response Time: <40 seconds Relative Humidity Range: 15 to 90% non-condensing Typical Baseline Range (pure air): -30 to +100ppm equivalent Maximum Zero Shift (+20°C to +40°C): -100ppm equivalent Long Term Output Drift: <2% signal loss/month Recommended Load Resistor: 10? Bias Voltage: Not required Repeatability: 1% of signal Output Linearity: Linear Colour Coding: Red
PHYSICAL CHARACTERISTICS Weight: 12.5 g nominal Position Sensitivity: None Storage Life: Six months in CTL container Recommended Storage Temp: 0°C to 20°C